If you are seeing or hearing this, then either your browser/user agent does not support Cascading Style Sheets, or you have elected to turn them off.  You may see and/or hear immediately following this notice alerts regarding national security or the Amber Alert system that do not apply.  To see if these alerts apply, follow the links provided.  -Commonwealth of Kentucky
The national Homeland Security Advisory System's Threat Condition is currently red, meaning severe.  Check your local radio or television stations for up-to-date information.  You may also check the National Homeland Security website or the Kentucky Homeland Security website for more information.  -Commonwealth of Kentucky
The national Homeland Security Advisory System's Threat Condition is currently orange, meaning high.  Check your local radio or television stations for up-to-date information, or you may visit the National Homeland Security website or the Kentucky Homeland Security website for more information.  -Commonwealth of Kentucky
An Amber Alert has been issued.  Visit the Kentucky Emergency Management website for information.  -Commonwealth of Kentucky
Logo for the Kentucky.gov site.  By clicking this logo, you will be taken to the Kentucky state home page. KY Agencies| KY Services  
www.water.ky.gov LogoAgency Name Image
Kentucky Unbridled Spirit-External Site
Get Adobe Reader
 
 
Last Modified:  12/6/2007
Turbidimeter

Protocol for Calibrating Turbidimeters
 
To effectively and accurately measure the clarity of water and provide an important “window” through which you can reliably judge the effectiveness of the treatment process, a properly calibrated turbidimeter is essential. 
 
Turbidimeters need to be calibrated on a quarterly basis or as often as recommended by the manufacturer. Check the owner’s manual and follow the manufacturer’s instructions for calibration and record the calibration readings in a logbook.

For online, continuous recording turbidimeters, refer to the technical guidance document from the Drinking Water Branch listed below:
 
 

These guidelines became effective September 2000 and should be incorporated into your Standard Operating Procedures (SOP) manual.

TWO WAYS TO CHECK TUBIDIMETER ACCURACY

1.  Calibration by comparison to a known standard. Use this method for quarterly checks.

2.  Calibration by comparison to another turbidimeter. Use this method on a monthly basis to verify the performance of online reporting meters. Verification checks assume that the meter being used for comparison has, itself, been calibrated to a known standard.  Ultimately, the validity of any turbidity reading depends upon the accuracy of the standard used to calibrate the turbidimeter.

TURBIDITY STANDARDS BACKGROUND

A turbidity standard is a liquid sample or other material having a defined and reproducible turbidity value. There are two types of turbidity calibration standards:

  • Primary standards are used to either directly calibrate a meter or to calibrate a secondary standard. Primary standards are the “gold standard” for turbidity. All calibrations must eventually be traceable to a primary standard. The most common acceptable form of primary standard is formazin. The other is a commercially manufactured liquid suspension of styrene divinyl benzene polymer beads (SDB). Either is acceptable because it will give repeatable results. Formazin may be easily produced in the lab or may be purchased in various prepared forms. While a 4,000 NTU stock solution may be good for a year, more dilute solutions used in calibration generally must be discarded after one week. SDB solutions are more costly but remain stable almost indefinitely.
  • Secondary standards include common examples such as liquid latex, SDB or Gelex solutions in a sealed sample container. They may also include glass rods or plates, plastic cylinders and mirror devices tailored for use in a specific manufacturer’s device. These standards are convenient, easy to use, cheap and reliable for daily calibration checks. However, to meet monitoring requirements, every secondary standard must first be calibrated to a primary standard. Also, since the secondary standard is likely to degrade with time, it is important to recheck its value against a primary standard every four months.

CALIBRATION TIPS

1.  Allow the turbidimeter to warm up for at least 30 minutes prior to use.

2.  Always use clean glassware and handle cuvettes so there are no fingerprints in the area where light passes through the sample.

3.  Use alcohol and a lint-free cloth to clean the outside of the cuvettes.

4.  Once the cuvette becomes scratched or pitted, replace with a new cuvette.

5.  Rinse cuvette with about 0.5 ml of the standard to be used. Discard. Repeat with a second rinse to ensure that only the standard is present in the cuvette.

6.  DO NOT pour used primary standard back into the bottle. This contaminates the contents.

7.  Allow the standard to reach room temperature before using to alleviate fogging of the cuvette.

8.  If problems are encountered with calibration, allow the sample to sit loosely capped with a clean cap until all air bubbles have settled out or gently tap sides of cuvette to release air bubbles, then take another reading.

If you have questions about calibration, using turbidimeters or your Standard Operating Procedure (SOP) manual, speak with technical ssistance staff listed below: 

  • Julie Roney, supervisor of Technical Assistance, phone 502-564-3410
  • Jim Hamon, Frankfort, phone 502-564-3410
  • Eric Sutton, Frankfort, phone 502-564-3410
  • Crystal Davis, Columbia Regional Office, phone 270-384-4734
  • Don DeKoster, Louisville Regional Office, phone 502-429-7122
  • Jackie Logsdon, Madisonville Regional Office, phone 270-824-7529
DEP Division of Water
14 Reilly Road
Frankfort, KY 40601
Phone: 502-564-3410
Fax: 502-564-0111
E-mail: water@ky.gov